Radiation Dosimetry of Dental Enamel Using X-Band and Q-Band EPR Spectroscopy
ORAL
Abstract
Electron paramagnetic resonance (EPR) dosimetry of tooth enamel can be used for individual dose reconstruction following radiation accidents. The purpose of this study was to develop a rapid, minimally invasive technique for obtaining a sample of dental enamel small enough to not disturb the structure and functionality of a tooth and to improve the sensitivity of the spectral signals using X-band (9.4 GHz) and Q-band (34 GHz) EPR spectroscopy. EPR measurements in X-band were performed on 100 mg isotropic powdered enamel samples and Q-band measurements done on 4 mg (1x1x3 mm) enamel biopsy samples. All samples were obtained from discarded teeth collected during normal dental treatment. In order to study the variation of the Radiation-Induced Signal (RIS) at different orientations in the applied magnetic field samples were placed in the resonance cavity for Q-band EPR. In X-band spectra, the RIS is distinct from the ``native'' radiation-independent signal only for doses $>$ 0.5Gy. Q-band, however, resolves the RIS and ``native'' signals and improves sensitivity by a factor of 20 enabling measurements in 2-4 mg tooth enamel samples.
–
Authors
-
Tania De
Howard University
-
Alex Romanyukha
Naval Dosimetry Center
-
Barry Pass
Howard University
-
Prabhakar Misra
Howard University