Improved Spectral Fitting Models for the \hbox{B-Stark} Diagnostic at DIII-D

POSTER

Abstract

Recent results are presented from the B-Stark diagnostic installed on the \hbox{DIII-D} tokamak. This diagnostic provides measurements of the magnitude and direction of the internal magnetic field. The \hbox{B-Stark} system is a version of a motional Stark effect (MSE) diagnostic based on the relative line intensities and spacing of the Stark split $D_\alpha$ emission from injected neutral beams. Improvements to the spectral fitting model are presented, including the addition of an analytical model for $D_\alpha$ emission from the fast-ion distribution. We discuss the accuracy of using in-situ beam-into-gas calibrations to find the beam emission line profiles, the viewing direction and the transmission properties of the collection optics. We also present results of efforts to improve the determination of the beam emission line profiles. Finally, the magnetic field measured with the \hbox{B-Stark} system is compared to values found from plasma equilibrium reconstructions (EFIT) and the MSE polarimetry system on \hbox{DIII-D}.

*Supported by the US DOE under DE-FG02-07ER54917, DE-AC02-09CH11466, DE-FC02-04ER54698, and DE-AC52-07NA27344.