A New UV Spectrometer for Measurements of Tungsten Erosion {\&} Re-deposition

POSTER

Abstract

A new high-resolution and high-throughput ultraviolet spectrometer has been constructed to resolve the most promising W line radiation arising from erosion during W sourcing experiments in the DIII-D divertor. Spectral surveys in the Compact Toroidal Hybrid experiment between 200 and 400 nm have identified over 60 neutral and singly ionized W emission lines in the UV region, which can be combined with atomic predictions to determine the net erosion and re-deposition of W on plasma facing surfaces. The importance of W metastable level populations requires that multiple W emission lines be monitored simultaneously to accurately characterize erosion rates. The new UV spectrometer has a maximum resolving power of 1.6 {\AA} at 250 nm with better than 1 kHz temporal resolution. The details of the fiber-coupled collection optics and instrument shielding required for installation on the DIII-D tokamak will be presented along with expected signal levels for DIII-D plasma conditions.

*Work supported by USDOE grants DE-SC0015877, DE-FG02-00ER54610 & DE-FC02-04ER54698.