Calibration of Optically-Stimulated Luminescence Dose Measurements for X-Ray Fluorescence Bone Lead Measurements
ORAL
Abstract
An experimental phantom-based method was established in our lab for lead (Pb) concentration measurements in the human tibia bone using a microbeam-based L-shell x-ray fluorescence (LXRF). \textit{In vivo} LXRF bone lead measurements, however, require knowledge of the radiation dose delivered by the microbeam to the soft tissue overlying the bone. Optically-stimulated luminescence dosimeters (OSLDs) were selected for such dose measurements due to their reduced size similar to the small irradiated tissue volumes. Available calibrations of commercial OSLDs mostly target clinical applications such as x-ray imaging modalities and radiation therapy and could not be used for LXRF bone Pb measurements due to large differences in the x-ray photon energy. OSLD calibration was performed using an ion chamber calibrated by an x-ray beam with similar quality factor to the microbeam. Calculations and exposure rate measurements were performed at six positions along the microbeam. Calculated values were larger than the measured ones by a factor in the 1.2 to 1.7 range. Differences were explained by calculations not accounting the backscattered x-rays and the varying x-ray beam size.
–
Authors
-
Sarah Kroeker
California State University, Fresno
-
MIHAI GHERASE
Physics, California State University, Fresno