Adsorbed Hydrogen Film Densities and Thicknesses Determined from Low-Temperature Hydrogen Sorption Experiments

ORAL

Abstract

Hydrogen storage through physisorption has shown tremendous promise. Advancement of our understanding about hydrogen behavior in confined pores can lead to a development of new storage materials. For example, isosteric heat is used to determine the quality of a sorbent. Yet, Clausius-Clapeyron isosteric heat calculations are typically performed on excess adsorption, which leads to unphysical results. Absolute adsorption should be used for these calculations. To determine absolute adsorption from excess adsorption, the volume of the adsorbed film is needed. We have built a Sievert type instrument capable of temperatures from 10 K to 300 K and pressures up to 200 bar. Using this instrument to measure low temperature ($<$ 77 K) and high pressure ($>$ 100 bar) isotherms, experimental film density and volume have been determined from the linear decrease in excess H$_{2}$ as a function of bulk gas density. Additionally, some materials have shown H$_{2}$ uptakes higher than what their surface area predicts. One hypothesis is N$_{2}$, the standard gas to determine surface areas, is sterically forbidden to go into pores that H$_{2}$ can. Sub-critical H$_{2}$ isotherms have been measured to determine surface area available to the H$_{2}$ and comparisons are made to N$_{2}$ surface area.

Authors

  • Jacob Burress

    Physics Dept. U. of Missouri

  • Elmar Dohnke

    Physics Dept. U. of Missouri

  • Matthew Beckner

    Physics Dept. U. of Missouri

  • Mark Lee

    Chemistry Dept. U. of Missouri

  • Carlos Wexler

    Physics Dept. U. of Missouri

  • Peter Pfeifer

    Physics Dept. U. of Missouri